CR-3 CRYOGENIC HYGROMETER

 

The new CR-3, like the CR-2, utilizes a closed cycle cryocooler, but differs in that it cannot be used on aircraft. The CR-3 is specifically designed for long-term, continuous-use applications where the instrument will mounted in a fixed location such as a laboratory. 

Cryocooler lifetime is greatly extended for the CR-3. Typically, MTBF is over 100,000 hours for the cooler.

Options include a special sample chamber and mirror assembly for toxic and corrosive gases. Like all our instruments, the CR-3 is helium-leak checked. In addition to voltage, current and alarm outputs, the CR-3 has an RS-232 port.

To download a brochure detailing the CR-3 specifications, click here.

To download the Operating Manual for the CR-3 Model, click here.

CR-3 Hygrometer Specifications

Measurement Range
  Dew / frost point temp.    °C -120 to +30°C
  Lowest Concentration ppbv  
    sea level  <1
    100 psig (7.8 bar)  <.5
Reading Accuracy    °C
  Entire Range ±0.1°C
  Concentration ±2%
Typical Response Time 10-20 seconds
Operating Range
  Temperature °C -40 to +40°C
  Pressure, psia (bar) 5  to 115 psia (0.3 to 7.8 bar)
  Air / Gas flow range  liters/min. 0.2 to 3 liters/min
Cooling Mechanism Kleemenko Cycle Cryocooler
Power Requirements
Standard 115 vac, 60 Hz, <250 W
Optional 220-240 vac, 50-60Hz
Outputs 0-10V, 4-20mA, Alarm, RS-232